PicoVNA Vector Network Analyzer
The Impressive PicoVNA: Low Cost, High Performance
PicoVNA 106 and PicoVNA 108 are compact, cost-effective, USB-controlled Vector Network Analyzers (VNAs) offering bandwidths of up to 8.5 GHz, delivering performance that far exceeds their price point.
Accurate: With a dynamic range of up to 124 dB at the highest resolution bandwidth, and RMS trace noise as low as 0.005 dB, the PicoVNA ensures that your measurements accurately reflect the true characteristics of the Device Under Test (DUT).
Fast: Perform up to 5,500 dual-port S-parameter measurements per second. A 201-point, two-port .s2p file can be generated in less than 38 ms.
Affordable: The instrument delivers exceptional value, while its calibration kits are also cost-effective and easy to maintain, helping keep the overall cost of ownership low.
Reliable: A four-receiver architecture minimizes uncorrectable errors and measurement latency.
Simple: Automatic calibration with an E-Cal kit makes setup quick and easy, allowing you to get up and running with minimal effort.
| Mode | PicoVNA 106 | PicoVNA 108 | Conditions |
| Measurement bandwidth | 140 kHz、70 kHz、35 kHz、15 kHz、10 kHz、5 kHz、1 kHz、500 Hz、100 Hz、50 Hz、10 Hz | ||
| Average displayed noise floor | Band(MHz) Typical(dB) Max(dB) | Band(MHz) Typical(dB) Max(dB) 0.3 to 1 -100 -90 1 to 6000 -124 -110 > 6000 -120 -100 | Relative to the test signal level set to maximum power after an S21 calibration. Ports terminated as during the isolation calibration step. |
| Dynamic range (click for graphs) | 0.3 MHz to 10 MHz 10 MHz to 6 GHz | 0.3 MHz to 10 MHz 0.3 MHz to 8.5 GHz | 10 Hz bandwidth Maximum test power: PicoVNA 106: +6 dBm PicoVNA 108: 0 dBm No averaging |
| Temperature stability, typical | 0.02 dB/°C for F < 4 GHz 0.04 dB/°C for F ≥ 4 GHz | Measured after an S21 calibration | |
| Trace noise, dB RMS | Band Typical Max | 201-point sweep covering 1 MHz to 6 GHz or 8.5 GHz. | |
| Spurious responses | -76 dBc( typical),-70 dBc(Max)。 | The main spurious response occurs at close to (2 x RF + 1.3) or (3 x RF + 2.6) MHz, where RF is the test frequency in MHz. For example, when testing a bandpass filter with a centre frequency of, say 1900 MHz, an unwanted response will occur around 632.47 or 949.35 MHz. In all known cases the levels will be as stated. | |